The PFC175SFX is a 175W SFX power supply specifically designed for powering microATX medical applications. The PFC175SFX exceeds the SFX12V version 3.0 standard including PCI Express.
The PFC175SFX is a 175W SFX power supply specifically designed for powering microATX medical applications. The PFC175SFX exceeds the SFX12V version 3.0 standard including PCI Express.
Factor Correction insures conformance to EN61000-3-2 harmonic standard and EN61000-4-11 Voltage Dips and Interruptions.
Cable connectors and lengths are customized to your specifications. Please consult our output cable configurator or contact us for details.
ATX12VO Option: The PFC175SFX can be built to conform to the Intel 12-Volt only (ATX12VO) configuration, using the 10-pin motherboard power connector.
Recommended Line Cord: SJT, 3X16 or 3X14 AWG
Input Voltage Range: 100 to 250 VAC +/-10%
Supply overvoltage: Cat Two <300V
Input current: 2.5 A MAX
Frequency: 50/60 HZ
Inrush Current: 45A Cold @ 264 VAC
Efficiency: 75% to 80% TYP.
PFC Correction: Active > .95
Leakage Current: < 250uA @ 264 VAC 60HZ
Input Fuses: 2X 5A 250V Line & Neutral
Max output power: 175 Watts cont.
Hold-up Time: >40ms @ full load
Overvoltage Protection: +15% of set voltage
Thermal Protection: Auto Recovery/ cycle pwr. on
Minimum Loading: None
Cross regulation: <0.5%
Line Regulation: <0.05%
Load Regulation: <1% +/- cable factor
AC turn on time: < 2 sec
Output Voltage | +12V | +5V | +3.3V | -12 | -5V | +5Vsb |
Output current max. | 10A |
15A | 15A | 1.0A |
0.5A | 2.0A |
Ripple mV P-P | 100 | 50 | 50 | 100 | 50 | 50 |
Operational Temperature: 0 to 50 C
Storage Temperature: -40 to 80 C
Operational Humidity: 0 to 95%NC
Storage Humidity: 0 to 95%NC
Operational Altitude: -500 to 9840 FT
Storage Altitude: -500 to 40.000 FT
Pollution Degree Class: Two
MTBF (demonstrated): > 500,000 hours
Expected Life: >8 years when use as specified
EN61000-4-2 ESD level3
EN61000-4-3 RF susceptibility 3v/m
EN61000-4-4 EFT level3
EN61000-4-5 Surge Level3
EN61000-3-2 Harmonics
EN61000-3-3 Flicker
EN61000-4-11 Voltage Dips, Short Interruptions
EN55011 Conducted and radiated emissions
Desktop Platform Form Factors: REV 1.1 March 07 IEC60601-1 2nd & 3rd edition
RoHS compliant UL file E191947 ISO9001:2008
Lead free ROHS and REACH compliant
ECCN# EAR99